First Report of Extraspinal Lead Migration Along a Thoracic Spinal Nerve After Spinal Cord Stimulation.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 32540296.
- Also identified by DOI 10.1016/j.wneu.2020.06.062.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Spinal cord stimulation for failed back surgery syndrome and chronic pain is a well-established treatment regimen today. Lead migration is the most common complication; mainly epidural caudal more than cranial electrode migration from the primary position is described repeatedly throughout the literature. A 60-year-old male patient with failed back surgery syndrome was eligible for spinal cord stimulation. Surgery had been performed 4 weeks before readmission with proper lead positioning of both electrodes in the midline of the epidural space. The electrode fixation mechanism at L2/3 had to be revised and was replaced with multiple ligature fixations due to the patient's slim build. He presented to our outpatient clinic with thoracic right-sided pain matching T5 with signs of overstimulation of the paravertebral muscles. X-ray imaging revealed cranial migration of 1 lead to T4 and a right-sided extraspinal migration of the other lead along a spinal nerve in T5 exiting the neuroforamen and following beneath the corresponding rib dorsally. Revision surgery was performed using a thoracic paddle electrode. Lead migration remains a challenge in spinal cord stimulation regardless of the fixation method. Rare unusual migration patterns in addition to simple caudal or cranial migration might pose a challenge for revision surgery and thus might reduce overall treatment efficacy.
Medical subject headings
- Epidural Space
- Failed Back Surgery Syndrome
- Spinal Cord Stimulation
- Thoracic Nerves