Magnetic resonance cholangiopancreatography with compressed sensing at 1.5 T: clinical application for the evaluation of branch duct IPMN of the pancreas.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 32556465.
- Also identified by DOI 10.1007/s00330-020-06996-2 and PMC identifier 7554004.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
To evaluate magnetic resonance cholangiopancreatography (MRCP) with compressed sensing (CS) for the assessment of branch duct intraductal papillary mucinous neoplasm (BD-IPMN) of the pancreas. For this purpose, conventional navigator-triggered (NT) sampling perfection with application-optimized contrast using different flip angle evolutions (SPACE) MRCP was compared with various CS-SPACE-MRCP sequences in a clinical setting. A total of 41 patients (14 male, 27 female, mean age 68 years) underwent 1.5-T MRCP for the evaluation of BD-IPMN. The MRCP protocol consisted of the following sequences: conventional NT-SPACE-MRCP, CS-SPACE-MRCP with long (BHL, 17 s) and short single breath-hold (BHS, 8 s), and NT-CS-SPACE-MRCP. Two board-certified radiologists evaluated image quality, duct sharpness, duct visualization, lesion conspicuity, confidence, and communication with the main pancreatic duct in consensus using a 5-point scale (1-5), with higher scores indicating better quality/delineation/confidence. Maximum intensity projection reconstructions and originally acquired data were used for evaluation. Wilcoxon signed-rank test was used to compare the intra-individual difference between sequences. BHS-CS-SPACE-MRCP had the highest scores for image quality (3.85 ± 0.79), duct sharpness (3.81 ± 1.05), and duct visualization (3.81 ± 1.01). There was a significant difference compared with NT-CS-SPACE-MRCP (p < 0.05) but no significant difference to the standard NT-SPACE-MRCP (p > 0.05). Concerning diagnostic quality, BHS-CS-SPACE-MRCP had the highest scores in lesion conspicuity (3.95 ± 0.92), confidence (4.12 ± 1.08), and communication (3.8 ± 1.06), significantly higher compared with NT-SPACE-MRCP, BHL-SPACE-MRCP, and NT-CS-SPACE-MRCP (p = <0.05). MRCP with CS 3D SPACE for the evaluation of BD-IPMN at 1.5 T provides the best results using a short breath-hold sequence. This approach is feasible and an excellent alternative to standard NT 3D MRCP sequences. • 1.5-T MRCP with compressed sensing for the evaluation of branch duct IPMN is a feasible method. • Short breath-hold sequences provide the best results for this purpose.
Medical subject headings
- Cholangiopancreatography, Magnetic Resonance
- Pancreatic Intraductal Neoplasms
- Pancreatic Neoplasms