Retroauricular thyroidectomy with a single-arm robotic surgical system: Preclinical cadaveric study.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 32852084.
- Also identified by DOI 10.1002/hed.26436.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In this study, we evaluate a new flexible, single-arm robotic system in performing thyroidectomy via a retroauricular approach. Four fresh cadavers were used. Technical details of the new system are discussed, and the comprehensive surgical procedures are described step by step. This single-port flexible system could be used to successfully perform retroauricular thyroidectomy. The ideal angle to dock the patient-side cart was at a 90° angle to the operating table. When the cannula tip was placed 10 to 15 cm away from the skin incision, positioning and full movement of all four instruments without collisions were possible. Three flexible instruments and an endoscope made the robotic dissection more efficient. We describe the first preclinical evaluation of an innovative, flexible, single-arm robotic surgical system for retroauricular thyroidectomy.