Biomechanical Analysis of an Interspinous Process Fixation Device with In Situ Shortening Capabilities: Does Spinous Process Compression Improve Segmental Stability?
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 32891838.
- Also identified by DOI 10.1016/j.wneu.2020.08.203.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The objective of this study was to characterize the biomechanical implications of spinous process compression, via in situ shortening of a next-generation interspinous process fixation (ISPF) device, in the context of segmental fusion. Seven lumbar cadaveric spines (L1-L4) were tested. Specimens were first tested in an intact state, followed by iterative instrumentation at L2-3 and subsequent testing. The order followed was 1) stand-alone ISPF (neutral height); 2) stand-alone ISPF (shortened in situ from neutral height; shortened); 3) lateral lumbar interbody fusion (LLIF) + ISPF (neutral); 4) LLIF + ISPF (shortened); 5) LLIF + unilateral pedicle screw fixation; 6) LLIF + bilateral pedicle screw fixation. A 7.5-Nm moment was applied in flexion/extension, lateral bending, and axial rotation via a kinematic test frame. Segmental range of motion (ROM) and lordosis were measured for all constructs. Comparative analysis was performed. Statistically significant flexion/extension ROM reductions: all constructs versus intact condition (P < 0.01); LLIF + ISPF (neutral and shortened) versus stand-alone ISPF (neutral and shortened) (P < 0.01); LLIF + USPF versus ISPF (neutral) (P = 0.049); bilateral pedicle screw fixation (BPSF) versus stand-alone ISPF (neutral and shortened) (P < 0.01); LLIF + BPSF versus LLIF + unilateral pedicle screw fixation (UPSF) (P < 0.01). Significant lateral bending ROM reductions: LLIF + ISPF (neutral and shortened) versus intact condition and stand-alone ISPF (neutral) (P < 0.01); LLIF + UPSF versus intact condition and stand-alone ISPF (neutral and shortened) (P < 0.01); LLIF + BPSF versus intact condition and all constructs (P < 0.01). Significant axial rotation ROM reductions: LLIF + ISPF (shortened) and LLIF + UPSF versus intact condition and stand-alone ISPF (neutral) (P ≤ 0.01); LLIF + BPSF versus intact condition and all constructs (P ≤ 0.04). In situ shortening of an adjustable ISPF device may support increased segmental stabilization compared with static ISPF.
Medical subject headings
- Internal Fixators
- Joint Instability