Metal artifacts of hip arthroplasty implants at 1.5-T and 3.0-T: a closer look into the B<sub>1</sub> effects.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 32918566.
- Also identified by DOI 10.1007/s00256-020-03597-4.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To evaluate the effect of circular polarization (CP) and elliptical polarization (EP) of the B<sub>1</sub> field on metal implant-induced artifacts of titanium (Ti) and cobalt-chromium (CoCr) hip arthroplasty implants at 1.5-T and 3.0-T field strengths. In vitro Ti and CoCr total hip arthroplasty implants were evaluated using high transmit and receive bandwidth turbo spin echo (HBW-TSE) and slice encoding for metal artifact correction (SEMAC) metal artifact reduction techniques. Each technique was implemented at 1.5-T, which only allows for CP of B<sub>1</sub> field as the system default, as well as 3.0-T, which permitted CP and EP. Manual segmentation quantified the size of the metal artifacts at the level of the acetabular cup, femoral neck, and femoral shaft. In the acetabular cup and femoral neck, 1.5-T CP achieved smaller artifact sizes than 3.0-T CP (28-29% on HBW-TSE, p = 0.002-0.005; 17-34% on SEMAC, p = 0.019-0.102) and 3.0-T EP (25-28% on HBW-TSE, p = 0.010-0.011; 14-36% on SEMAC, p = 0.058-0.135) techniques. In the femoral stem region, 3.0-T EP achieved more efficient artifact suppression than 3.0-T CP (HBW-TSE 44-45%, p < 0.001-0.022; SEMAC 76-104%, p < 0.001-0.022) and 1.5-T CP (HBW-TSE 76-96%, p < 0.001-0.003; SEMAC 138-173%, p = 0.003-0.005) techniques. Despite slightly superior metal reduction ability of the 1.5-T in the region of the acetabular cup and prosthesis neck, 3.0-T MRI of hip arthroplasty implants using elliptically polarized RF pulses may overall be more effective in reducing metal artifacts than the current standard 1.5-T MRI techniques, which by default implements circularly polarized RF pulses.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Artifacts
Anatomy
- hip