In vitro methodology for medical device material thrombogenicity assessments: A use condition and bioanalytical proof-of-concept approach.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 32929881.
- Also identified by DOI 10.1002/jbm.b.34705 and PMC identifier 7821245.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
Device manufacturers and regulatory agencies currently utilize expensive and often inconclusive in vivo vascular implant models to assess implant material thrombogenicity. We report an in vitro thrombogenicity assessment methodology where test materials (polyethylene, Elasthane™ 80A polyurethane, Pebax®), alongside positive (borosilicate glass) and negative (no material) controls, were exposed to fresh human blood, with attention to common blood-contact use conditions and the variables: material (M), material surface modification (SM) with heparin, model (Mo), time (T), blood donor (D), exposure ratio (ER; cm<sup>2</sup> material/ml blood), heparin anticoagulation (H), and blood draw/fill technique (DT). Two models were used: (1) a gentle-agitation test tube model and (2) a pulsatile flow closed-loop model. Thrombogenicity measurements included thrombin generation (thrombin-antithrombin complex [TAT] and human prothrombin fragment F1.2), platelet activation (β-thromboglobulin), and platelet counts. We report that: (a) thrombogenicity was strongly dependent (p < .0001) on M, H, and T, and variably dependent (p < .0001 - > .05) on Mo, SM, and D (b) differences between positive control, test, and negative control materials became less pronounced as H increased from 0.6 to 2.0 U/ml, and (c) in vitro-to-in vivo case comparisons showed consistency in thrombogenicity rankings on materials classified to be of low, moderate, and high concern. In vitro methods using fresh human blood are therefore scientifically sound and cost effective compared to in vivo methods for screening intravascular materials and devices for thrombogenicity.
Medical subject headings
- Biocompatible Materials
- Blood Coagulation
- Blood Platelets
- Materials Testing
- Platelet Activation
- Thrombosis