Safety and accuracy of robot-assisted placement of pedicle screws compared to conventional free-hand technique: a systematic review and meta-analysis.
meta_analysis · Level I
Where this comes from
- Record sourced from PubMed, PMID 32976997.
- Also identified by DOI 10.1016/j.spinee.2020.09.007.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The introduction and integration of robot technology into modern spine surgery provides surgeons with millimeter accuracy for pedicle screw placement. Coupled with computer-based navigation platforms, robot-assisted spine surgery utilizes augmented reality to potentially improve the safety profile of instrumentation. In this study, the authors seek to determine the safety and efficacy of robotic-assisted pedicle screw placement compared to conventional free-hand (FH) technique. We conducted a systematic review of the electronic databases using different MeSH terms from 1980 to 2020. The present study measures pedicle screw accuracy, complication rates, proximal-facet joint violation, intraoperative radiation time, radiation dosage, and length of surgery. A total of 1,525 patients (7,379 pedicle screws) from 19 studies with 777 patients (51.0% with 3,684 pedicle screws) in the robotic-assisted group were included. Perfect pedicle screw accuracy, as categorized by Gerztbein-Robbin Grade A, was significantly superior with robotic-assisted surgery compared to FH-technique (Odds ratio [OR]: 1.68, 95% confidence interval [CI]: 1.20-2.35; p=.003). Similarly, clinically acceptable pedicle screw accuracy (Grade A+B) was significantly higher with robotic-assisted surgery versus FH-technique (OR: 1.54, 95% CI: 1.01-2.37; p=.05). Furthermore, the complication rates and proximal-facet joint violation were 69% (OR: 0.31, 95% CI: 0.20-0.48; p<.00001) and 92% less likely (OR: 0.08, 95% CI: 0.03-0.20; p<.00001) with robotic-assisted surgery versus FH-group. Robotic-assisted pedicle screw implantation significantly reduced intraoperative radiation time (MD: -5.30, 95% CI: -6.83-3.76; p<.00001) and radiation dosage (MD: -3.70, 95% CI: -4.80-2.60; p<.00001) compared to the conventional FH-group. However, the length of surgery was significantly higher with robotic-assisted surgery (MD: 22.70, 95% CI: 6.57-38.83; p=.006) compared to the FH-group. This meta-analysis corroborates the accuracy of robot-assisted pedicle screw placement.
Medical subject headings
- Pedicle Screws
- Robotic Surgical Procedures
- Robotics
- Spinal Fusion
- Surgery, Computer-Assisted
- Zygapophyseal Joint