Automated Embryo Manipulation and Rotation via Robotic nDEP-Tweezers.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 33052848.
- Also identified by DOI 10.1109/TBME.2020.3031043.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Embryo manipulation is a fundamental task in assisted reproductive technology (ART). Nevertheless, conventional pick-place techniques often require proper alignment to avoid causing damage to the embryo and further, the tools have limited capability to orient the embryo being handled. This paper presents a novel and non-invasive technique that can easily manipulate mouse embryos on a polyvinyl chloride (PVC) Petri dish. An inverted microchip with quadrupole electrodes was attached to a micromanipulator to become a robotic dielectrophoresis (DEP) tweezers, and a motorized platform provided additional mobility to the embryos lying on a Petri dish. Vision-based algorithms were developed to evaluate relevant information of the embryos from the image, and to provide feedback signals for precise position and orientation control of the embryo. A series of experiments was conducted to examine the system performance, and the embryo can be successfully manipulated to a specified location with the desired orientation for subsequent processing. This system offers a non-contact, low cost, and flexible method for rapid cell handling. As the DEP tweezers can grasp the embryo without the need for precise alignment, the overall time required to process a large number of embryos can be shortened.
Medical subject headings
- Robotic Surgical Procedures