Single-center midterm results with the low-profile Zenith Alpha thoracic endovascular stent graft.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 33065242.
- Also identified by DOI 10.1016/j.jvs.2020.10.003.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The analysis of endovascular treatment of thoracic aortic diseases using new low-profile stent grafts in large series is crucial to understanding the durability of these devices. The present study reports the midterm outcomes of a single-center experience using the Zenith Alpha thoracic endovascular stent graft. The outcomes of 270 procedures performed on 262 patients (197 men; mean age, 70.5 ± 9.5 years) using the Zenith Alpha thoracic endovascular stent graft from November 2013 to December 2019 for different thoracic aortic diseases were analyzed. The primary endpoints were 30-day clinical success and midterm (5-year) clinical success. The secondary endpoints were the adverse event rate at 30 days and midterm and access- and device-related complications. The follow-up of surviving patients was performed using computed tomography angiography and office visits at 1, 6, and 12 months and annually thereafter. Kaplan-Meier analysis was performed for overall survival, and freedom from thoracic aortic endovascular repair-related mortality and related reinterventions. The overall 30-day mortality was 5.2% (2.5% for elective and 30.8% for nonelective cases). Type I endoleaks were identified in six patients. The 30-day primary technical and clinical success rates were 97.8% and 92.6%, respectively. Femoral cutdown was used in 41.1% of cases and percutaneous access in 58.5%. The rate of femoral artery complications after the percutaneous approach was 5.1%, with the need for surgical conversion in 1.9%. The stroke rate was 4.1% (major stroke, 2.2%), and the spinal cord ischemia rate was 3.7% (permanent paraplegia, 0.7%). Of the 248 survivors, 239 complied with the follow-up protocol with adequate computed tomography angiograms available images. Overall follow-up survival was 94.0% at 1 year, 91.6% at 2 years, 88.9% at 3 years, and 88.5% at 5 years. The unplanned secondary endovascular procedure rate was 5.3%. No stent fractures or new-onset type I endoleaks due to stent graft migration were observed in the study cohort. The midterm outcomes of this new generation of low-profile devices were satisfactory. The reported low incidence of secondary procedures and the absence of migrations are promising for the long-term durability of these devices.
Medical subject headings
- Aorta, Thoracic
- Aortic Aneurysm, Thoracic
- Blood Vessel Prosthesis
- Blood Vessel Prosthesis Implantation
- Diverticulum
- Endovascular Procedures
- Stents