Emergency Splicing of Transected Ventricular Assist Device Driveline.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 33075317.
- Also identified by DOI 10.1016/j.athoracsur.2020.07.073.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Left ventricular assist device (LVAD) driveline transection is an often fatal complication traditionally treated with salvage surgery. We present the case of a 70-year-old man who presented in cardiogenic shock after accidentally transecting his LVAD driveline. The driveline was emergently dissected at the bedside, all 6 internal wires were spliced, and his LVAD was restarted without morbidity. He subsequently underwent formal durable driveline splice by the device manufacturer 8 hours later and was discharged home uneventfully 28 hours after presentation. In this report we describe our novel, rapid, reproducible method for emergency HeartWare HVAD (Medtronic, Framingham, MA) driveline splicing, offering an alternative to surgical device exchange.
Medical subject headings
- Device Removal
- Heart Failure
- Heart-Assist Devices