Modification of a domiciliary ventilator to increase FiO<sub>2</sub>: an off-label modification which may be of value in COVID-19.
review · Level V
Where this comes from
- Record sourced from PubMed, PMID 33077616.
- Also identified by DOI 10.1136/thoraxjnl-2020-215487.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Although nasal continuous positive airway pressure or non-invasive ventilation is used to manage some patients with acute lung injury due to COVID-19, such patients also demonstrate increased minute ventilation which makes it hard, if the device is used in line with the manufacturer's instructions, to achieve adequate oxygen delivery. In addition, if a hospital contains many such patients, then it is possible that the oxygen requirements will exceed infrastructure capacity. Here we describe a simple modification of two exemplar ventilators normally used for domiciliary ventilation, which substantially increased the fraction of inspired oxygen (FiO<sub>2</sub>) delivered.