The implementation of a rapid sample preparation method for the detection of SARS-CoV-2 in a diagnostic laboratory in South Africa.
Where this comes from
- Record sourced from PubMed, PMID 33079951.
- Also identified by DOI 10.1371/journal.pone.0241029 and PMC identifier 7575110.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The SARS-CoV-2 pandemic has resulted in shortages of both critical reagents for nucleic acid purification and highly trained staff as supply chains are strained by high demand, public health measures and frequent quarantining and isolation of staff. This created the need for alternate workflows with limited reliance on specialised reagents, equipment and staff. We present here the validation and implementation of such a workflow for preparing samples for downstream SARS-CoV-2 RT-PCR using liquid handling robots. The rapid sample preparation technique evaluated, which included sample centrifugation and heating prior to RT-PCR, showed a 97.37% (95% CI: 92.55-99.28%) positive percent agreement and 97.30% (95% CI: 90.67-99.52%) negative percent agreement compared to nucleic acid purification-based testing. This method was subsequently adopted as the primary sample preparation method in the Groote Schuur Hospital Virology Diagnostic Laboratory in Cape Town, South Africa.
Medical subject headings
- Betacoronavirus
- Clinical Laboratory Techniques
- Coronavirus Infections
- Laboratories, Hospital
- Pneumonia, Viral
- Reverse Transcriptase Polymerase Chain Reaction
- Robotics