Change in knee biomechanics during squat and walking induced by a modification in TKA size.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 33093756.
- Also identified by DOI 10.1016/j.jor.2020.10.006 and PMC identifier 7559522.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study is to analyze the effects of TKA under-dimensioning during daily activities. A regular ("control") size and an undersized design of the same fixed bearing asymmetric PS prosthesis were analyzed during walking and squat using finite element analysis. The two models showed similar internal-external rotations and antero-posterior displacements during both activities. Slightly higher displacements, wider contact areas and lower contact pressure were found in the control size. Post-cam engagement angles were similar on both sizes. Changes in TKA size slightly affected knee kinematics and kinetics, with post-cam related differences leading to minor changes in kinetic patterns.