Transmission of Oscillatory Volumes into the Preterm Lung during Noninvasive High-Frequency Ventilation.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 33095994.
- Also identified by DOI 10.1164/rccm.202007-2701OC.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
<b>Rationale:</b> There is increasing evidence for a clinical benefit of noninvasive high-frequency oscillatory ventilation (nHFOV) in preterm infants. However, it is still unknown whether the generated oscillations are effectively transmitted to the alveoli.<b>Objectives:</b> To assess magnitude and regional distribution of oscillatory volumes (V<sub>Osc</sub>) at the lung level.<b>Methods:</b> In 30 prone preterm infants enrolled in a randomized crossover trial comparing nHFOV with nasal continuous positive airway pressure, electrical impedance tomography recordings were performed. During nHFOV, the smallest amplitude to achieve visible chest wall vibration was used, and the frequency was set at 8 hertz.<b>Measurements and Main Results:</b> Thirty consecutive breaths during artifact-free tidal ventilation were extracted for each of the 228 electrical impedance tomography recordings. After application of corresponding frequency filters, Vt and V<sub>Osc</sub> were calculated. There was a signal at 8 and 16 Hz during nHFOV, which was not detectable during nasal continuous positive airway pressure, corresponding to the set oscillatory frequency and its second harmonic. During nHFOV, the mean (SD) V<sub>Osc</sub>/Vt ratio was 0.20 (0.13). Oscillations were more likely to be transmitted to the non-gravity-dependent (mean difference [95% confidence interval], 0.041 [0.025-0.058]; <i>P</i> < 0.001) and right-sided lung (mean difference [95% confidence interval], 0.040 [0.019-0.061]; <i>P</i> < 0.001) when compared with spontaneous Vt.<b>Conclusions:</b> In preterm infants, V<sub>Osc</sub> during nHFOV are transmitted to the lung. Compared with the regional distribution of tidal breaths, oscillations preferentially reach the right and non-gravity-dependent lung. These data increase our understanding of the physiological processes underpinning nHFOV and may lead to further refinement of this novel technique.
Medical subject headings
- Continuous Positive Airway Pressure
- Forced Expiratory Volume
- High-Frequency Ventilation
- Infant, Premature
- Noninvasive Ventilation
- Respiratory Distress Syndrome, Newborn