Cardiac output estimation using pulse wave analysis-physiology, algorithms, and technologies: a narrative review.
review · Level V
Where this comes from
- Record sourced from PubMed, PMID 33246581.
- Also identified by DOI 10.1016/j.bja.2020.09.049.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Pulse wave analysis (PWA) allows estimation of cardiac output (CO) based on continuous analysis of the arterial blood pressure (AP) waveform. We describe the physiology of the AP waveform, basic principles of PWA algorithms for CO estimation, and PWA technologies available for clinical practice. The AP waveform is a complex physiological signal that is determined by interplay of left ventricular stroke volume, systemic vascular resistance, and vascular compliance. Numerous PWA algorithms are available to estimate CO, including Windkessel models, long time interval or multi-beat analysis, pulse power analysis, or the pressure recording analytical method. Invasive, minimally-invasive, and noninvasive PWA monitoring systems can be classified according to the method they use to calibrate estimated CO values in externally calibrated systems, internally calibrated systems, and uncalibrated systems.
Medical subject headings
- Cardiac Output
- Monitoring, Physiologic
- Pulse Wave Analysis