Hemolysis induced by Left Ventricular Assist Device is associated with proximal tubulopathy.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 33253314.
- Also identified by DOI 10.1371/journal.pone.0242931 and PMC identifier 7703997.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Chronic subclinical hemolysis is frequent in patients implanted with Left Ventricular Assist Device (LVAD) and is associated with adverse outcomes. Consequences of LVADs-induced subclinical hemolysis on kidney structure and function is currently unknown. Thirty-three patients implanted with a Heartmate II LVAD (Abbott, Inc, Chicago IL) were retrospectively studied. Hemolysis, Acute Kidney Injury (AKI) and the evolution of estimated Glomerular Filtration Rate were analyzed. Proximal Tubulopathy (PT) groups were defined according to proteinuria, normoglycemic glycosuria, and electrolytic disorders. The Receiver Operating Characteristic (ROC) curve was used to analyze threshold of LDH values associated with PT. Median LDH between PT groups were statistically different, 688 IU/L [642-703] and 356 IU/L [320-494] in the "PT" and "no PT" groups, respectively p = 0.006. To determine PT group, LDH threshold > 600 IU/L was associated with a sensitivity of 85.7% (95% CI, 42.1-99.6) and a specificity of 84.6% (95% CI, 65.1-95.6). The ROC's Area Under Curve was 0.83 (95% CI, 0.68-0.98). In the "PT" group, patients had 4.2 [2.5-5.0] AKI episodes per year of exposure, versus 1.6 [0.4-3.7] in the "no PT" group, p = 0.03. A higher occurrence of AKI was associated with subsequent development of Chronic Kidney Disease (CKD) (p = 0.02) and death (p = 0.05). LVADs-induced subclinical hemolysis is associated with proximal tubular functional alterations, which in turn contribute to the occurrence of AKI and subsequent CKD. Owing to renal toxicity of hemolysis, measures to reduce subclinical hemolysis intensity as canula position or pump parameters should be systematically considered, as well as specific nephroprotective therapies.
Medical subject headings
- Acute Kidney Injury
- Fanconi Syndrome
- Heart Failure
- Heart-Assist Devices