Sufficient sampling for kriging prediction of cortical potential in rat, monkey, and human µECoG.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 33326943.
- Also identified by DOI 10.1088/1741-2552/abd460 and PMC identifier 8058280.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
<i>Objective</i>. Large channel count surface-based electrophysiology arrays (e.g. µECoG) are high-throughput neural interfaces with good chronic stability. Electrode spacing remains ad hoc due to redundancy and nonstationarity of field dynamics. Here, we establish a criterion for electrode spacing based on the expected accuracy of predicting unsampled field potential from sampled sites.<i>Approach</i>. We applied spatial covariance modeling and field prediction techniques based on geospatial kriging to quantify sufficient sampling for thousands of 500 ms µECoG snapshots in human, monkey, and rat. We calculated a probably approximately correct (PAC) spacing based on kriging that would be required to predict µECoG fields at≤10% error for most cases (95% of observations).<i>Main results</i>. Kriging theory accurately explained the competing effects of electrode density and noise on predicting field potential. Across five frequency bands from 4-7 to 75-300 Hz, PAC spacing was sub-millimeter for auditory cortex in anesthetized and awake rats, and posterior superior temporal gyrus in anesthetized human. At 75-300 Hz, sub-millimeter PAC spacing was required in all species and cortical areas.<i>Significance</i>. PAC spacing accounted for the effect of signal-to-noise on prediction quality and was sensitive to the full distribution of non-stationary covariance states. Our results show that µECoG arrays should sample at sub-millimeter resolution for applications in diverse cortical areas and for noise resilience.
Medical subject headings
- Auditory Cortex
- Electrocorticography