Continuous Glucose Monitoring in the Intensive Care Unit During the COVID-19 Pandemic.

Agarwal, Shivani; Mathew, Justin; Davis, Georgia M; Shephardson, Alethea; Levine, Ann; Louard, Rita; Urrutia, Agustina; Perez-Guzman, Citlalli et al. · Diabetes Care · 2021

case_series · Level IV

Where this comes from

Abstract

Real-time continuous glucose monitoring (rtCGM) in critically ill hospitalized patients holds promise; however, real-world data are needed. We placed Dexcom G6 CGM on intensive care unit (ICU) patients at Montefiore Medical Center with confirmed coronavirus disease 2019 (COVID-19) infection and glycemic variability. We analyzed inpatient CGM accuracy using point-of-care (POC) glucose-CGM matched pairs and included patients for analysis regardless of clinical status. We included 11 patients with CGM: 8 on continuous insulin infusion (CII), 8 on vasopressors, 8 intubated, 4 on high-dose glucocorticoids, 6 on renal replacement therapy, and 2 with anasarca. Accuracy was 12.58% for mean and 6.3% for median absolute relative difference. CGM reduced POC testing by ∼60% for patients on CII. In this real-world preliminary analysis of rtCGM during critical illness, we demonstrate early feasibility, considerable accuracy, and meaningful reduction in the frequency of POC glucose testing.