Fully Closed Loop Glucose Control With a Bihormonal Artificial Pancreas in Adults With Type 1 Diabetes: An Outpatient, Randomized, Crossover Trial.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 33397767.
- Also identified by DOI 10.2337/dc20-2106.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To demonstrate the performance and safety of a bihormonal (insulin and glucagon) artificial pancreas (AP) in adults with type 1 diabetes. In this outpatient, randomized, crossover trial, 2-week fully closed loop glucose control (AP therapy) was compared with 2-week open loop control (patient's normal insulin pump therapy with a glucose sensor if they had one). A total of 23 patients were included in the analysis. Time in range (70-180 mg/dL [3.9-10 mmol/L]) was significantly higher during closed loop (median 86.6% of time [interquartile range 84.9-88.5]) compared with open loop (53.9% [49.7-67.2]; <i>P</i> < 0.0001). Compared with insulin pump therapy, the bihormonal AP provided superior glucose control, without meal or exercise announcements, and was safe in adults with type 1 diabetes.
Medical subject headings
- Diabetes Mellitus, Type 1
- Pancreas, Artificial