First in human experience with an epicardial beating heart device for secondary mitral regurgitation.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 33436291.
- Also identified by DOI 10.1016/j.jtcvs.2020.11.169.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We describe a novel, off-pump, epicardial implant that is intended to reshape both the mitral valve annulus and the left ventricle (LV) in those with secondary mitral regurgitation (MR). Five patients underwent an epicardial implant with the Mitral Touch device (Mitre Medical Corp, Morgan Hill, Calif), during concomitant off-pump coronary artery bypass for secondary MR. The median age was 71.2 years; 4 patients had severe MR and 1 moderate. Patients were followed for 1 year with transthoracic echocardiography and computed tomography. Safety, cardiac remodeling, and MR were assessed by an independent core laboratory. One patient died within 30 days from nondevice-related organ failure and the remaining 4 survived through 1-year follow-up. Implant technical success was 100% and took an average of 52 minutes. Paired computed tomography showed mean left ventricular end-systolic volume remodeling at 1 and 12 months of -35% and -31%, respectively. They averaged left atrial end-systolic volume remodeling of -12% and -15% at 1 and 12 months. Right ventricular end-systolic volume changes of -19% and -8% and right atrial end-systolic volume remodeling of -5% and 1%, at the 1- and 12-month time points were noted. Regurgitant volume by transthoracic echocardiography decreased by 46% and 44% and the ejection fraction from 34.6% to 32.1% and 39.5%, at 1 and 12 months, respectively. There were no device-related complications reported to 1 year. The Epicardial Mitral Touch System for Mitral Regurgitation (ENRAPT-MR) study demonstrates a first-in-man, off-pump, epicardial repair of secondary MR. Procedural safety and geometric correction of the mitral valve apparatus and LV was achieved. Further studies in the United States are underway.
Medical subject headings
- Heart Valve Prosthesis
- Heart Valve Prosthesis Implantation
- Heart-Assist Devices
- Hemodynamics
- Mitral Valve
- Mitral Valve Annuloplasty
- Mitral Valve Insufficiency
- Ventricular Function, Left