Melody Valve Fracture Causing Mitral Stenosis: Novel Solution for Transapical Valve-in-Valve.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 33497660.
- Also identified by DOI 10.1016/j.athoracsur.2020.12.051.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The use of Melody valves in the mitral position has been introduced in clinical practice. Stent fracture is a recognized complication of Melody valve implantation in the pulmonary position; however, reports in the mitral position are rare. We present the case of an 8-year-old boy in whom complete fracture of the proximal stent struts occurred, causing acute severe mitral stenosis, and in whom urgent hybrid transapical Melody valve implantation in the fractured Melody valve was performed successfully using a novel modified technique. This modification simplified the implantation, led to reduced time of the procedure, and minimized hemodynamic instability.
Medical subject headings
- Heart Valve Prosthesis
- Heart Valve Prosthesis Implantation
- Mitral Valve Stenosis
- Prosthesis Failure