Quantitative Assessment of Viral Dispersion Associated with Respiratory Support Devices in a Simulated Critical Care Environment.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 33534659.
- Also identified by DOI 10.1164/rccm.202008-3070OC and PMC identifier 8314904.
- Licence recorded as CC BY-NC-ND.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
<b>Rationale:</b> Patients with severe coronavirus disease (COVID-19) require supplemental oxygen and ventilatory support. It is unclear whether some respiratory support devices may increase the dispersion of infectious bioaerosols and thereby place healthcare workers at increased risk of infection with severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2).<b>Objectives:</b> To quantitatively compare viral dispersion from invasive and noninvasive respiratory support modalities.<b>Methods:</b> This study used a simulated ICU room with a breathing-patient simulator exhaling nebulized bacteriophages from the lower respiratory tract with various respiratory support modalities: invasive ventilation (through an endotracheal tube with an inflated cuff connected to a mechanical ventilator), helmet ventilation with a positive end-expiratory pressure (PEEP) valve, noninvasive bilevel positive-pressure ventilation, nonrebreather face masks, high-flow nasal oxygen (HFNO), and nasal prongs.<b>Measurements and Main Results:</b> Invasive ventilation and helmet ventilation with a PEEP valve were associated with the lowest bacteriophage concentrations in the air, and HFNO and nasal prongs were associated with the highest concentrations. At the intubating position, bacteriophage concentrations associated with HFNO (2.66 × 10<sup>4</sup> plaque-forming units [PFU]/L of air sampled), nasal prongs (1.60 × 10<sup>4</sup> PFU/L of air sampled), nonrebreather face masks (7.87 × 10<sup>2</sup> PFU/L of air sampled), and bilevel positive airway pressure (1.91 × 10<sup>2</sup> PFU/L of air sampled) were significantly higher than those associated with invasive ventilation (<i>P</i> < 0.05 for each). The difference between bacteriophage concentrations associated with helmet ventilation with a PEEP valve (4.29 × 10<sup>-1</sup> PFU/L of air sampled) and bacteriophage concentrations associated with invasive ventilation was not statistically significant.<b>Conclusions:</b> These findings highlight the potential differential risk of dispersing virus among respiratory support devices and the importance of appropriate infection prevention and control practices and personal protective equipment for healthcare workers when caring for patients with transmissible respiratory viral infections such as SARS-CoV-2.
Medical subject headings
- Critical Care
- DNA, Viral
- Disease Transmission, Infectious
- Respiratory Insufficiency
- Ventilators, Mechanical
- Virus Diseases
- Viruses