Biomedical Photoacoustic Imaging With Unknown Spatially Distributed Ultrasound Sensor Array.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 33534699.
- Also identified by DOI 10.1109/TBME.2021.3056715.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
With the growth of interest in different medical study on biological function, non-invasive photoacoustic imaging of biological tissue attracts the interests for researchers. To eliminate the limited angle effect of photoacoustic imaging based on ultrasound linear array, spatially distributed ultrasound sensor array is applied. The accurate sensor array position determines the quality of the imaging results. In this study, we proposed three methods based on photoacoustic and ultrasound signals to enhance the imaging quality using a 256-element full-ring array. Groups of photoacoustic and ultrasound signals are used to regress the position of each element sensor. In phantom study and mouse brain study, photoacoustic imaging results can both yield details clearly with average error rate of less than 1% (50 [Formula: see text]). The performance of our three methods have proved that they can be potentially applied to other ultrasound-based medical imaging studies with unknown distributed positions of sensor array to enhance the imaging quality. The proposed methods can contribute to precise biomedical imaging with unknown distributed positions of sensor array in different application scenarios.
Medical subject headings
- Photoacoustic Techniques