A Unique Mode of Failure in the Noncontact Bridging Periprosthetic Plating System.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 33543880.
- Also identified by DOI 10.5435/JAAOSGlobal-D-20-00143 and PMC identifier 7861960.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Although lateral locking plates are often a preferred and successful fixation construct for the treatment of periprosthetic proximal and distal femur fractures, specific complications and modes of failure have been associated and well-described with their use. We present two cases of implant failure in the Non-Contact Bridge Periprosthetic Plating System (Zimmer Biomet) in which a nonlocked screw fretted through the annular seating of the plate. One case demonstrates failure in the setting of a proximal femur periprosthetic fracture, whereas the other demonstrates failure in the setting of a distal femur periprosthetic fracture. This unique mode of failure has not previously been reported in the literature.
Medical subject headings
- Femoral Fractures
- Periprosthetic Fractures
Anatomy
- femur