Automatic dosimetric verification of online adapted plans on the Unity MR-Linac using 3D EPID dosimetry.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 33582193.
- Also identified by DOI 10.1016/j.radonc.2021.01.037.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The Unity MR-Linac is equipped with an EPID, the images from which contain information about the dose delivered to the patient. The purpose of this study was to introduce a framework for the automatic dosimetric verification of online adapted plans using 3D EPID dosimetry and to present the obtained dosimetric results. The framework was active during the delivery of 1207 online adapted plans corresponding to 127 clinical IMRT treatments (74 prostate, 19 rectum, 19 liver and 15 lymph node oligometastases). EPID reconstructed dose distributions in the patient geometry were calculated automatically and then compared to the dose distributions calculated online by the treatment planning system (TPS). The comparison was performed by γ-analysis (3% global/2mm/10% threshold) and by the difference in median dose to the high-dose volume (ΔHDV<sub>D50</sub>). 85% for γ-pass rate and 5% for ΔHDV<sub>D50</sub> were used as tolerance limit values. 93% of the online plans were verified automatically by the framework. Missing EPID data was the reason for automation failure. 91% of the verified plans were within tolerance. Automatic dosimetric verification of online adapted plans on the Unity MR-Linac is feasible using in vivo 3D EPID dosimetry. Almost all online adapted plans were approved automatically by the framework. This newly developed framework is a major step forward towards the clinical implementation of a permanent safety net for the entire online adaptive workflow.
Medical subject headings
- Radiotherapy Planning, Computer-Assisted
- Radiotherapy, Intensity-Modulated