Adjustable breathing resistance for laryngectomized patients: Proof of principle in a novel heat and moisture exchanger cassette.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 33615610.
- Also identified by DOI 10.1002/hed.26571 and PMC identifier 8048959.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
Due to the heat and moisture exchanger's (HME) breathing resistance, laryngectomized patients cannot always use an (optimal) HME during physical exercise. We propose a novel HME cassette concept with adjustable "bypass," to provide adjustment between different breathing resistances within one device. Under standardized conditions, the resistance and humidification performance of a high resistance/high humidification HME (XM) foam in a cassette with and without bypass were compared to a lower resistance/lesser humidification HME (XF) foam in a closed cassette. With a bypass in the cassette, the resistance and humidification performance of XM foam were similar to those of XF foam in the closed cassette. Compared to XM foam in the closed cassette, introducing the bypass resulted in a 40% resistance decrease, whereas humidification performance was maintained at 80% of the original value. This HME cassette prototype allows adjustment between substantially different resistances while maintaining appropriate humidification performances.
Medical subject headings
- Hot Temperature
- Laryngectomy