Transcatheter Heart Valve Downstream Fluid Dynamics in an Accelerated Evaluation Environment.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 33638029.
- Also identified by DOI 10.1007/s10439-021-02751-w.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Transcatheter aortic valve replacements (TAVRs) provide minimally invasive delivery of bioprosthetic heart valves (BHVs) for the treatment of aortic valve disease. While surgical BHVs show efficacy for 8-10 years, long-term TAVR durability remains unknown. Pre-clinical testing evaluates BHV durability in an ISO:5840 compliant accelerated wear tester (AWT), yet, the design and development of AWTs and their accuracy in predicting in vivo performance, is unclear. As a result of limited knowledge on AWT environment and BHV loading, durability assessment of candidate valves remains fundamentally empirical. For the first time, high-speed particle image velocimetry quantified an ISO:5840 compliant downstream AWT velocity field, Reynolds stresses, and turbulence intensity. TAVR enface imaging quantified the orifice area and estimated the flow rate. When valve area and flow rate were at their maximum during peak systole (1.49 cm<sup>2</sup> and 16.05 L/min, respectively), central jet velocity, Reynolds normal and shear stress, and turbulence intensity grew to 0.50 m/s, 265.1 Pa, 124.6 Pa, and 37.3%, respectively. During diastole, unique AWT recirculation produced retrograde flow and the directional changes created eddies. These novel AWT findings demonstrated a substantially reduced valve fully loaded period and pressure not matching in vivo or in vitro studies, despite the comparable fluid environment and TAVR motion.
Medical subject headings
- Bioprosthesis
- Heart Valve Prosthesis
- Materials Testing