Deformation mechanics of self-expanding venous stents: Modelling and experiments.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 33730560.
- Also identified by DOI 10.1016/j.jbiomech.2021.110333.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Deformation properties of venous stents based on braided design, chevron design, Z design, and diamond design are compared using in vitro experiments coupled with analytical and finite element modelling. Their suitability for deployment in different clinical contexts is assessed based on their deformation characteristics. Self-expanding stainless steel stents possess superior collapse resistance compared to Nitinol stents. Consequently, they may be more reliable to treat diseases like May-Thurner syndrome in which resistance against a concentrated (pinching) force applied on the stent is needed to prevent collapse. Braided design applies a larger radial pressure particularly for vessels of diameter smaller than 75% of its nominal diameter, making it suitable for a long lesion with high recoil. Z design has the least foreshortening, which aids in accurate deployment. Nitinol stents are more compliant than their stainless steel counterparts, which indicates their suitability in veins. The semi-analytical method presented can aid in rapid assessment of topology governed deformation characteristics of stents and their design optimization.
Medical subject headings
- Alloys
- Stents