Performance evaluation of bactericidal effect and endotoxin inactivation by low-temperature ozone/hydrogen peroxide mixed gas exposure.

Nomura, Yusuke; Yamamura, Junji; Fukui, Chie; Fujimaki, Hideo; Sakamoto, Kazuyuki; Matsuo, Ken-Ichi; Kuromatsu, Hisashi; Kikuchi, Yutaka et al. · J Biomed Mater Res B Appl Biomater · 2021

basic_science · Level V

Where this comes from

Abstract

This study evaluated the performance of a new O<sub>3</sub> /H<sub>2</sub> O<sub>2</sub> mixed gas sterilization instrument for killing microorganisms and inactivating bacterial endotoxin at low temperatures. Sterility assurance level was achieved by an over 6-log reduction of Geobacillus stearothermophilus ATCC 12980, and the decimal reduction value was 0.77 min in sterilization mode. A reduction of over 3 logs in Limulus amebocyte lysate coagulation activity of purified endotoxin from Escherichia coli was observed after treatment in endotoxin-inactivation mode. The same inactivation ability was observed when treating dried bacterial cells. Biomaterials made of polymer or metal did not exhibit cytotoxicity after gas exposure at O<sub>3</sub> concentrations below 200 ppm. As the results of human cell-based pyrogen testing, significant amounts of endotoxin that were over the limit for medical devices contacting cerebrospinal fluid (2.15 EU/device) were detected on scissors washed with a washer-disinfector and sterilized with ethylene oxide or autoclaving. In contrast, endotoxin decreased to 0.29 ± 0.05 EU/device after O<sub>3</sub> /H<sub>2</sub> O<sub>2</sub> mixed gas sterilization in endotoxin-inactivation mode. Compared to conventional gas sterilization methods, O<sub>3</sub> /H<sub>2</sub> O<sub>2</sub> mixed gas has high sterilization ability and a strong capacity to inactivate endotoxin. It is expected that this sterilization technology will improve the safety of reusable medical devices and utensils for regenerative medicine.

Medical subject headings