Efficient and Fast Immuno-Labeling of Clarified Tissues Using Low-Field Enhanced Diffusion.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 33788676.
- Also identified by DOI 10.1109/TBME.2021.3070146.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To alleviate the severe limitation of the prohibitively long process of immune-fluorescence labeling on the routine applications of revolutionary intact tissue clearing techniques in diverse biomedical arenas. We proposed an easily adaptable approach, electro-enhanced rapid staining (EERS), for highly efficient and fast immuno-labeling of thick clarified tissues. In EERS, an optimized and precisely controlled weak external electric field is engineered into a compact device to enable efficient and uniform transport of antibodies into clarified tissues while minimizing the detrimental effect of macromolecular crowding at the tissue-solution interface. The experimental results show that, with EERS, a current density of only ∼0.2 mA mm<sup>-2</sup> is sufficient to achieve uniform labeling of clarified tissues of several millimeters thick in a few hours without detectable tissue damage. In addition, the amount of antibodies required is also several-fold lower than conventional immuno-labeling assays under comparable conditions. It is expected that the implementation of EERS in most laboratories should significantly expedite the application of tissue clearing in a broad range of research explorations, both basic and clinical.
Medical subject headings
- Antibodies