Reproducibility and accuracy of a target motion mitigation technique for dose-escalated prostate stereotactic body radiotherapy.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 33992627.
- Also identified by DOI 10.1016/j.radonc.2021.05.004.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To quantitate the accuracy, reproducibility and prostate motion mitigation efficacy rendered by a target immobilization method used in an intermediate-risk prostate cancer dose-escalated 5×9Gy SBRT study. An air-inflated (150 cm<sup>3</sup>) endorectal balloon and Foley catheter with three electromagnetic beacon transponders (EBT) were used to mitigate and track intra-fractional target motion. A 2 mm margin was used for PTV expansion, reduced to 0 mm at the interface with critical OARs. EBT-detected ≥ 2 mm/5 s motions mandated treatment interruption and target realignment prior to completion of planned dose delivery. Geometrical uncertainties were measured with an in-house ESAPI script. Quantitative data were obtained in 886 sessions from 189 patients. Mean PTV dose was 45.8 ± 0.4 Gy (D95 = 40.5 ± 0.4 Gy). A mean of 3.7 ± 1.7 CBCTs were acquired to reach reference position. Mean treatment time was 19.5 ± 12 min, 14.1 ± 11 and 5.4 ± 5.9 min for preparation and treatment delivery, respectively. Target motion of 0, 1-2 and >2 mm/10 min were observed in 59%, 30% and 11% of sessions, respectively. Temporary beam-on hold occurred in 7.4% of sessions, while in 6% a new reference CBCT was required to correct deviations. Hence, all sessions were completed with application of the planned dose. Treatment preparation time > 15 min was significantly associated with the need of a second reference CBCT. Overall systematic and random geometrical errors were in the order of 1 mm. The prostate immobilization technique explored here affords excellent accuracy and reproducibility, enabling normal tissue dose sculpting with tight PTV margins.
Medical subject headings
- Prostatic Neoplasms
- Radiosurgery
- Radiotherapy, Intensity-Modulated