Effect of pH on the Quantification of Common Chemistry Analytes in Body Fluid Specimens Using the Roche cobas Analyzer for Clinical Diagnostic Testing.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 34075393.
- Also identified by DOI 10.1093/ajcp/aqab025.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To determine the influence of pH on recovery of analytes in body fluids (BFs), investigate the mechanism of pH interference, measure the frequency of abnormal-pH BFs received, and compare pH measured by meter and paper. We performed pH titration in residual BFs. A low-pH BF was spiked and neutralized to investigate pH interference. We measured analytes on a Roche cobas c501 analyzer (Roche Diagnostics) and calculated the percent recovery. Measurement of pH using a meter and paper was conducted on 122 BF samples received in the laboratory. Enzyme activity in BFs was unaffected when pH = 7.4-8.5 lactate dehydrogenase, pH = 7.3-10.2 amylase, pH = 6.0-9.9 lipase, and pH = 1.3-11.7 all other analytes. BFs had mean (range) pH of 8.0 (5.1-8.9), with a mean (range) difference (paper ‒ meter) of ‒0.4 (‒0.6 to 1.1). Irreversible loss of enzyme activity occurs in BFs at low pH. Few clinical BFs have pH < 7.0, but laboratories should incorporate pH measurement in BF workflows.
Medical subject headings
- Body Fluids
- Diagnostic Tests, Routine
- Enzyme Assays
- Hydrogen-Ion Concentration