Exhalation Spreading During Nasal High-Flow Therapy at Different Flow Rates.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 34135285.
- Also identified by DOI 10.1097/CCM.0000000000005009 and PMC identifier 8204857.
- Licence recorded as CC BY-NC-ND.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Severe acute respiratory syndrome coronavirus 2 is transmitted through aerosols and droplets. Nasal high-flow therapy could possibly increase the spreading of exhalates from patients. The aim of this study is to investigate whether nasal high-flow therapy affects the range of the expiratory plume compared with spontaneous breathing. Interventional experiment on single breaths of a healthy volunteer. Research laboratory at the Bauhaus-University Weimar. A male subject. Videos and images from a schlieren optical system were analyzed during spontaneous breathing and different nasal high-flow rates. The maximal exhalation spread was 0.99, 2.18, 2.92, and 4.1 m during spontaneous breathing, nasal high-flow of 20 L/min, nasal high-flow of 40 L/min, and nasal high-flow of 60 L/min, respectively. Spreading of the expiratory plume in the sagittal plane can completely be blocked with a surgical mask. Nasal high-flow therapy increases the range of the expiratory air up to more than 4 meters. The risk to pick up infectious particles could be increased within this range. Attachment of a surgical mask over the nasal high-flow cannula blocks the expiratory airstream.
Medical subject headings
- Cannula
- Exhalation
- Oxygen Inhalation Therapy
- Pulmonary Ventilation