Precision and accuracy of FEV1 measurements from the Vitalograph copd-6 mini-spirometer in a healthy Ugandan population.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 34181689.
- Also identified by DOI 10.1371/journal.pone.0253319 and PMC identifier 8238209.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Evaluate the accuracy and precision of the copd-6 mini-spirometer for FEV1 in a rural Ugandan population. In a cross-sectional study, 171 smallholder farmers performed spirometry with copd-6, and a diagnostic-quality spirometer. The copd-6 underestimated FEV1 at low flows and overestimated FEV1 at high flows. Across all participants, the device slightly overestimated FEV1 by 0.04 [0.02; 0.06] L. Calibration data showed similar patterns. The copd-6 could be considered as an affordable tool for research on lung function impairment in resource-constrained settings. However, further validation in a study population with obstructive lung disease is needed.