Paediatric headbox as aerosol and droplet barrier.
Where this comes from
- Record sourced from PubMed, PMID 34266877.
- Also identified by DOI 10.1136/archdischild-2020-321546.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
High-flow nasal oxygen (HFNO) is frequently used in hospitals, producing droplets and aerosols that could transmit SARS-CoV-2. To determine if a headbox could reduce droplet and aerosol transmission from patients requiring HFNO. The size and dispersion of propylene glycol (model for patient-derived infectious particles) was measured using a spectrometer and an infant mannequin receiving 10-50 L/min of HFNO using (1) no headbox, (2) open headbox, (3) headbox-blanket or (4) headbox with a high-efficiency particulate (HEP) filter covering the neck opening. All headbox set-ups reduced the dispersal of droplets and aerosols compared with no headbox. The headbox-blanket system increased aerosol dispersal compared with the open headbox. The fraction of aerosols retained in the headbox for HFNO of 10 and 50 L/min was, respectively, as follows: (1) open headbox: 82.4% and 42.2%; (2) headbox-blanket: 56.8% and 39.5%; (3) headbox-HEP filter: 99.9% and 99.9%. A HEP-filter modified headbox may serve as an effective droplet and aerosol barrier adjunct for the protection of staff caring for children receiving HFNO.