Magnetic Resonance-Guided Radiation Therapy of Patients With Cardiovascular Implantable Electronic Device on a 1.5 T Magnetic Resonance-Linac.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 34520872.
- Also identified by DOI 10.1016/j.prro.2021.08.011.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Magnetic resonance-guided radiation therapy is reported for treating patients with an insertable cardiac monitor and implantable cardiac pacemakers. All treatments were delivered using a 1.5 T MR-Linac. Among the 4 patients, 2 were treated with stereotactic body radiation therapy at a dose of 40 Gy in 5 fractions. A clinical safety protocol was developed to address the decision-making and patient selection, as well as the clarified responsibilities of different parties for management of patients with cardiovascular implantable electronic devices. Dose estimation based on out-of-field dose data are necessary for cardiovascular implantable electronic devices located outside the treatment fields.
Medical subject headings
- Defibrillators, Implantable
- Pacemaker, Artificial