Magnetic Resonance-Guided Radiation Therapy of Patients With Cardiovascular Implantable Electronic Device on a 1.5 T Magnetic Resonance-Linac.

Yang, Bin; Yuan, Jing; Cheung, Kin Yin; Huang, Chen-Yu; Poon, Darren M C; Yu, Siu Ki · Pract Radiat Oncol · 2022

case_series · Level IV

Where this comes from

Abstract

Magnetic resonance-guided radiation therapy is reported for treating patients with an insertable cardiac monitor and implantable cardiac pacemakers. All treatments were delivered using a 1.5 T MR-Linac. Among the 4 patients, 2 were treated with stereotactic body radiation therapy at a dose of 40 Gy in 5 fractions. A clinical safety protocol was developed to address the decision-making and patient selection, as well as the clarified responsibilities of different parties for management of patients with cardiovascular implantable electronic devices. Dose estimation based on out-of-field dose data are necessary for cardiovascular implantable electronic devices located outside the treatment fields.