Evaluation of the clinical performance of a magnetic force-assisted electrochemical immunoassay for the detection of SARS-CoV-2 antigens.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 34618868.
- Also identified by DOI 10.1371/journal.pone.0258394 and PMC identifier 8496795.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Rapid antigen (Ag) tests for the detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) provide quick results, do not require specialized technical skills or infrastructure, and can be used as a point-of-care method to prevent the spread of coronavirus disease (COVID-19). The performance of a magnetic force-assisted electrochemical immunoassay-based test, namely the MARK-B COVID-19 Ag test (BBB, Sungnam, Republic of Korea), was evaluated using 170 nasopharyngeal swab specimens and compared to that of RT-PCR and commercial rapid Ag test (STANDARD Q COVID-19 Ag Test, SD Biosensor, Suwon-si, Republic of Korea). The overall sensitivity and specificity of the MARK-B test were 90.0% (95% CI 79.4%-96.2%) and 99.0% (95% CI 95.0%-99.9%), respectively, with a kappa coefficient of 0.908. The correlations between the electrical current values of MARK-B and the Ct values of RT-PCR were -0.898 (E gene, 95% CI -0.938 to -0.834) and -0.914 (RdRp gene, 95% CI -0.948 to -0.860), respectively. The limit of detection of the MARK-B was measured using the viral culture reference samples and found to be 1 x 102 pfu/mL. The magnetic force-assisted electrochemical immunoassay-based Ag test can be used to rapidly detect SARS-CoV-2 infections, and the corresponding fully automated portable device can provide easy readability and semi-quantitative results.
Medical subject headings
- Antigens, Viral
- COVID-19
- COVID-19 Serological Testing
- Electrochemical Techniques
- Magnetic Fields
- SARS-CoV-2