Effects of Time Delay Between Unipolar Pulses in High Frequency Nano-Electrochemotherapy.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 34797759.
- Also identified by DOI 10.1109/TBME.2021.3129176.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
this work focuses on bleomycin electrochemotherapy using new modality of high repetition frequency unipolar nanosecond pulses. As a tumor model, Lewis lung carcinoma (LLC1) cell line in C57BL mice (n = 42) was used. Electrochemotherapy was performed with intertumoral injection of bleomycin (50 μL of 1500 IU solution) followed by nanosecond and microsecond range electrical pulse delivery via parallel plate electrodes. The 3.5 kV/cm pulses of 200 and 700 ns were delivered in a burst of 200 at frequencies of 1 kHz and 1 MHz. For comparison of treatment efficiency, a standard 1.3 kV/cm x 100 μs x 8 protocol was used. It was shown that it is possible to manipulate the efficacy of unipolar sub-microsecond electrochemotherapy solely by the time delay between the pulses. the results suggest that the sub-microsecond range pulses can be as effective as the protocols in European Standard Operating Procedures on Electrochemotherapy (ESOPE) using 100 μs pulses.
Medical subject headings
- Electrochemotherapy