Fundamentals of Camera-PPG Based Magnetic Resonance Imaging.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 34928810.
- Also identified by DOI 10.1109/JBHI.2021.3136603.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In Magnetic Resonance Imaging (MRI), cardiac triggering that synchronizes data acquisition with cardiac contractions is an essential technique for acquiring high-quality images. Triggering is typically based on the Electrocardiogram (ECG) signal (e.g. R-peak). Since ECG acquisition involves extra workflow steps like electrode placement and ECG signals are usually disturbed by magnetic fields in high Magnetic Resonance (MR) systems, we explored camera-based photoplethysmography (PPG) as an alternative. We used the in-bore camera of a clinical MR system to investigate the feasibility and challenges of camera-based cardiac triggering. Data from ECG, finger oximeter and camera were synchronously collected. Compared to finger-PPG, camera-based PPG provides a higher availability of the signal and the PPG marker delay relative to the ECG R-peak is considerably less with a camera monitoring the forehead. The insights obtained in this study provide a basis for an envisioned system-design phase.
Medical subject headings
- Electrocardiography
- Photoplethysmography