Biomechanical comparison of different fixation methods in metatarsal shaft fractures: A cadaver study.

Trost, Matthias; Knieps, Matthias; Hackl, Michael; Frey, Sönke; Scaal, Martin; Müller, Lars Peter; Dargel, Jens; Gick, Sascha · Clin Biomech (Bristol) · 2022

biomechanical · Level V

Where this comes from

Abstract

Various fixation methods are available for the operative treatment of metatarsal shaft fractures: Kirschner wire, Titanium elastic nail, plate, or an intramedullary bone stabilization system within a balloon catheter. The aim of this study was to compare the stability of these techniques. 72 metatarsals II to V from fresh frozen human cadaver feet were used. A shaft fracture was performed and fixed with a 1.6-mm Kirschner wire, a 1.5-mm Titanium elastic nail, a locking 6-hole-plate, or an intramedullary bone stabilization system. In a cantilever configuration, the head of the metatarsals was loaded statically (2 mm/min until failure; all groups) or cyclically (0 to 20 N for 1000 cycles with 10 mm/s, after 1000 cycles 2 mm/min until failure; plate and bone stabilization system). The mean failure strength for static loading was 17 N for Kirschner wire, 13 N for Titanium elastic nail, 73 N for plate and 34 N for the bone stabilization system (P < .01). For cyclic loading, the mean cycle of failure was 1000 for plate and 961 for the bone stabilization system (P = .76). The mean failure strength after cyclic loading was 73 N for plate and 48 N for the bone stabilization system (P = .03). Stability differs depending on the fixation method, with a plate showing the greatest stability and Kirschner wire or Titanium elastic nail the least. The stability of the bone stabilization system for fixing metatarsal shaft fractures is intermediate.

Medical subject headings

Anatomy