LDIAED: A lightweight deep learning algorithm implementable on automated external defibrillators.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 35213628.
- Also identified by DOI 10.1371/journal.pone.0264405 and PMC identifier 8880955.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Differentiating between shockable and non-shockable Electrocardiogram (ECG) signals would increase the success of resuscitation by the Automated External Defibrillators (AED). In this study, a Deep Neural Network (DNN) algorithm is used to distinguish 1.4-second segment shockable signals from non-shockable signals promptly. The proposed technique is frequency-independent and is trained with signals from diverse patients extracted from MIT-BIH, MIT-BIH Malignant Ventricular Ectopy Database (VFDB), and a database for ventricular tachyarrhythmia signals from Creighton University (CUDB) resulting, in an accuracy of 99.1%. Finally, the raspberry pi minicomputer is used to load the optimized version of the model on it. Testing the implemented model on the processor by unseen ECG signals resulted in an average latency of 0.845 seconds meeting the IEC 60601-2-4 requirements. According to the evaluated results, the proposed technique could be used by AED's.
Medical subject headings
- Databases, Factual
- Deep Learning
- Defibrillators
- Electrocardiography
- Ventricular Fibrillation