The validation of a low-cost inertial measurement unit system to quantify simple and complex upper-limb joint angles.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 35217243.
- Also identified by DOI 10.1016/j.jbiomech.2022.111000.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this research was to validate the use of a low-cost IMU system to measure upper-limb joint angles by comparing it to passive optical motion capture measures. Fifteen participants (five females; 25.9 ± 4.7 years) completed one trial of four simple range of motion (ROM) movements (elbow flexion/extension, shoulder abduction/adduction, shoulder flexion/extension, and shoulder internal/external rotation), and three complex functional daily tasks [hand to: back pocket (HBP), contralateral shoulder (HCS), head (HTH)]. Movements were measured, simultaneously, using fourteen OptiTrack cameras and five Notch® IMUs. The mean joint angle difference between devices ranged from 0.10° ± 3.11° for the HBP shoulder internal/external movement to 44.95° ± 3.50° for the simple ROM shoulder internal/external rotation movement. Nine of sixteen movement and plane comparisons showed significant differences between the device-specific movement cycle waveforms. Eleven of the comparisons showed either fixed and/or proportional biases (fixed only: 9; proportional only: 1; both fixed and proportional: 1). Due to multiple movements having large amplitude errors, low waveform similarities, and/or statistically significant mean differences between the Notch® IMUs and the gold-standard motion capture devices, we cannot recommend that Notch® IMUs are valid devices for measuring upper-limb joint angles during simple ROM and complex functional daily tasks.
Medical subject headings
- Shoulder Joint
- Upper Extremity