Abstract
Use of metal-on-metal (MoM) articulations in total hip arthroplasty (THA) has sharply declined due to high failure rates from metal-related complications. Although certain MoM designs have demonstrated only 46% survival, not all MoM designs have performed the same. The purpose of this study is to evaluate mid-term to long-term survival of a specific MoM implant with a modular titanium taper adapter. A retrospective review was performed on all patients who underwent primary THA at our center with the M<sup>2</sup>a-Magnum system (Zimmer Biomet, Warsaw, IN). Of 829 patients (956 hips) identified, 754 patients (869 hips) met inclusion criteria of signed research consent, minimum 2-year follow-up, and/or any revision surgery. Mean follow-up was 11.0 years (range 2-16; ±3.5). Mean cup angle of inclination was 42.8° (range 24°-70°, ±6.3°), with 88.0% reconstructed within the 40° ± 10° safe zone. There were 64 revisions (7.36%): 7 (0.81%) septic and 57 (6.56%) aseptic. Of those, 32 (3.68%) were adverse reactions to metal debris. Kaplan-Meier survival free of revision for all causes was 88.6% at 16 years (95% confidence interval 86.8-90.4). Univariate analysis of risk factors for all-cause, aseptic, and adverse reaction to metal debris revision found no relationship with female gender, age ≥65 years, body mass index >30 kg/m<sup>2</sup>, higher activity level, or inclination angle outlier. The results of this study demonstrate a more favorable mid-term to long-term survivorship with this specific MoM implant compared to other designs. Although our institution no longer performs MoM THA, further investigation into differences in MoM implant designs is warranted.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
- Metal-on-Metal Joint Prostheses
Anatomy
- hip