Air-Permeable Waterproofing Electrocardiogram Patch to Monitor Full-Day Activities for Multiple Days.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 35285162.
- Also identified by DOI 10.1002/adhm.202102703.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
On-skin healthcare patch-type devices have great technological challenges in monitoring full-day activities and wearing for multiple days without detachment. These challenges can be overcome when the sensor is air permeable but waterproof. This study presents a light-weight, highly stable, and stretchable Au electrode that is fabricated by sputtering on an imidized nanofiber mat. The contact surface of the electrode is hydro-wetting and the outer surface of the electrode is hydrophobic, so the porous electrode simultaneously has excellent sweat permeability and waterproofing capabilities. The electrode is applied to the electrocardiogram sensor for monitoring the cardiac signals for five consecutive days without detaching while doing various full-day activities such as relaxing, exercising, showering, and sleeping. This study suggests a modular setup of the electrodes and the cardiac signal processing unit for activating the device when cardiac monitoring is required.
Medical subject headings
- Wearable Electronic Devices