Efficacy of D-Wave Monitoring Combined With the Transcranial Motor-Evoked Potentials in High-Risk Spinal Surgery: A Retrospective Multicenter Study of the Monitoring Committee of the Japanese Society for Spine Surgery and Related Research.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 35343273.
- Also identified by DOI 10.1177/21925682221084649 and PMC identifier 10538305.
- Licence recorded as CC BY-NC-ND.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Retrospective multicenter cohort study. We aimed to clarify the efficacy of multimodal intraoperative neuromonitoring (IONM), especially in transcranial electrical stimulation of motor-evoked potentials (TES-MEPs) with spinal cord-evoked potentials after transcranial stimulation of the brain (D-wave) in the detection of reversible spinal cord injury in high-risk spinal surgery. We reviewed 1310 patients who underwent TES-MEPs during spinal surgery at 14 spine centers. We compared the monitoring results of TES-MEPs with D-wave vs TES-MEPs without D-wave in high-risk spinal surgery. There were 40 cases that used TES-MEPs with D-wave and 1270 cases that used TES-MEPs without D-wave. Before patients were matched, there were significant differences between groups in terms of sex and spinal disease category. Although there was no significant difference in the rescue rate between TES-MEPs with D-wave (2.0%) and TES-MEPs (2.5%), the false-positivity rate was significantly lower (0%) in the TES-MEPs-with-D-wave group. Using a one-to-one propensity score-matched analysis, 40 pairs of patients from the two groups were selected. Baseline characteristics did not significantly differ between the matched groups. In the score-matched analysis, one case (2.5%) in both groups was a case of rescue (<i>P</i> = 1), five (12.5%) cases in the TES-MEPs group were false positives, and there were no false positives in the TES-MEPs-with-D-wave group (<i>P</i> = .02). TES-MEPs with D-wave in high-risk spine surgeries did not affect rescue case rates. However, it helped reduce the false-positivity rate.