Choice of Spinal Interbody Fusion Cage Material and Design Influences Subsidence and Osseointegration Performance.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 35346883.
- Also identified by DOI 10.1016/j.wneu.2022.03.087.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The objective of the study was to quantify the effect of cage material (titanium-alloy vs. polyetheretherketone or PEEK) and design (porous vs. solid) on subsidence and osseointegration. Three lateral cages (solid PEEK, solid titanium, and 3-dimension-printed porous titanium cages) were evaluated for cage stiffness, subsidence compression stiffness, and dynamic subsidence displacement under simulated postoperative spine loading. Dowel-shaped implants made of grit-blasted solid titanium alloy (solid titanium) and porous titanium were fabricated using commercially available processes. Samples were processed for mechanical push-out testing and polymethylmethacrylate histology following an established ovine bone implantation model. The solid titanium cage exhibited the greatest stiffness (57.1 ± 0.6 kN/mm), followed by the porous titanium cage (40.4 ± 0.3 kN/mm) and the solid PEEK cage (37.1 ± 1.2 kN/mm). In the clinically relevant dynamic subsidence, the porous titanium cage showed the least amount of subsidence displacement (0.195 ± 0.012 mm), significantly less than that of the solid PEEK cage (0.328 ± 0.020 mm) and the solid titanium cage (0.538 ± 0.027 mm). Bony on-growth was noted histologically on all implant materials; however, only the porous titanium supported bony ingrowth with marked quantities of bone formed within the interconnected pores through 12 weeks. Functional differences in osseointegration were noted between groups during push-out testing. The porous titanium showed the highest maximum shear stress at 12 weeks and was the only group that demonstrated significant improvement (4-12 weeks). The choice of material and design is critical to cage mechanical and biological performances. A porous titanium cage can reduce subsidence risk and generate biological stability through bone on-growth and ingrowth.
Medical subject headings
- Osseointegration
- Spinal Fusion