Gated Tomographic Radionuclide Angiography Using 3D-Ring CZT StarGuide SPECT/CT Head-To-Head Comparison With a Cardiac-Dedicated CZT Camera: First Clinical Use and Validation.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 35353756.
- Also identified by DOI 10.1097/RLU.0000000000004153.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Gated tomographic radionuclide angiography can assess and monitor left cardiac function. Dedicated cardiac CZT cameras have enabled dose reduction and quicker acquisitions. New 3D-ring CZT general purpose systems are now available. We report 50 patients who underwent a 7-minute acquisition on a cardiac-dedicated CZT camera and 9 minutes on a new 3D-ring CZT system after mean injection of 321.4 ± 55.9 MBq 99mTc-labelled human serum albumin. There was no significant difference in left ventricular volumes, and left and right ventricular ejection fractions. These preliminary results seem to validate the use of 3D-ring CZT system for LEVF and cardiac function evaluation.
Medical subject headings
- Gamma Cameras
- Myocardial Perfusion Imaging