Fracture of the oxidized zirconium femoral component after total knee arthroplasty.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 35447466.
- Also identified by DOI 10.1016/j.knee.2022.03.014.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Femoral component fracture is a rare complication of total knee arthroplasty (TKA). We report a case of oxidized zirconium (Oxinium) femoral component fracture after total knee arthroplasty. The fracture site was the junction of the central and medial flanges. The patellar component and polyethylene insert had delamination at the contact point of the fracture line, and the tibial tray had loosening at the medial side. There was no cement adherence at the component fracture site, suggesting that debonding had occurred at the cement-implant interface in this area. Examination with a scanning electron microscope revealed beach marks, which are characteristic findings of metal fatigue. We considered that the cause of femoral component fracture was a fatigue fracture due to poor fixation of the component to the bone caused by poor osteotomy technique or poor cementing technique. To our knowledge, this is the first case of Oxinium femoral component fracture.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Femoral Fractures
- Knee Prosthesis
Anatomy
- femur
- knee