Templated freezing: a simple method may increase gripping force of the clamp on the tendon.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 35706018.
- Also identified by DOI 10.1186/s13018-022-03209-y and PMC identifier 9202137.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
To evaluate the effectiveness of combining a customized mold with frozen conventional clamps against other freezing and non-freezing methods. Forty-five porcine and 45 chicken tendons were evenly divided into five groups (n = 9 + 9/group): control group, non-freezing with gauze placed between tendon and clamp (gauze), non-freezing with suture fixation at tendon ends (suture), freezing with dry ice pocket placed at the clamps (pocket), and freezing using a templated liquid nitrogen clamp with a customized mold (mold). Tension tests were used to measure failure modes and loads. Slippage and avulsion were observed in non-freezing groups with significantly lower failure loads compared to freezing methods. With freezing, rupture occurred near the central point only in the mold group. The failure loads for porcine tendons in the mold group were higher (2121.651 ± 73.101 N) than the pocket group (1746.337 ± 68.849 N). The failure loads of chicken tendons in the mold (243.552 ± 15.881 N) and pocket groups (260.647 ± 22.161 N) were not statistically different. Freezing clamps represent the better choice for soft tissue clamping. The customized mold method could improve gripping effectiveness.
Medical subject headings
- Sutures
- Tendons