Distal Femoral Tracker Pin Placement Prevents Delayed Pin Tract-Induced Fracture in Robotic-Assisted Total Knee Arthroplasty: Results of Minimum 1-Year Follow-Up.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 35817054.
- Also identified by DOI 10.1055/s-0042-1749605.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to determine the outcomes of the placement of a pin tracker in the distal femur in robotic-assisted total knee arthroplasty (TKA) over a minimum follow-up period of 1 year. A consecutive series of 81 TKAs were performed in 59 patients using a robotic-assisted system (Mako) at our hospital. The mean follow-up period was 1.1 years (range, 1-1.2 years). No periprosthetic fracture at a pin tracker site was reported at a minimum of 1 year of follow-up. No patients reported minor or major complications associated with the distal femoral pins. In addition, unicortical pin-site drilling marks in all patients disappeared on one-postoperative-year follow-up radiographs. Our technique, which used unicortical pin placement in the distal femur in robotic-assisted TKA, demonstrated that it was a safe and reliable method at a minimum follow-up period of 1 year. The level of evidence is IV.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Robotic Surgical Procedures
- Femoral Fractures
Anatomy
- femur
- knee